How Much Forced Degradation in a Stress Study?

MTS Helpdesk

Question:

When I forcibly degrade my drug or drug product to create degradation products, how much degradation should I aim for?

Answer:

The aim of a forced degradation study is to generate degradation products from a drug which are both realistic and representative, typically for the purposes of assessing degradation pathways, or to create suitable samples that will allow the development of a suitable analytical method which can be used to assess the stability of the drug throughout its shelf-life.

If too little degradation occurs then it is difficult to evaluate what might happen in real time and to develop a method which can reliably determine the amount of the degradation products which could occur.

If too much degradation occurs then there is a risk that important degradation products may be further degraded and unimportant degradation products, which would not occur on storage, may be formed.

The generally accepted industry norm for the optimal amount of degradation is between 5 and 20%, typically aiming for around  10%. Anecdotally, there is a suggestion that the 10% corresponds to the usual specification for a drug product in the US of 90 to 110% label claim.

These numbers are not quoted in regulatory guidance with the exception of the guidance from Brazil (ANVISA Resolution RDC-53/2015) which directs that results must have > 10% degradation per condition and results that are < 10% degradation, after reasonable attempt, must have scientific justification.

(Note: This article was originally posted on LinkedIn)

Submit a Question to the MTS Helpdesk

Do you have a question for our expert relating to the topic of analytical chemistry? Use our enquiry form to send it to us.

What is an Analytical Procedure Lifecycle?
MTS Helpdesk Question: What are the elements of the Analytical Procedure Lifecycle, as mentioned in …
Limit of Quantitation: Does it Matter?
MTS Helpdesk Question: Is the limit of quanitation always important when validating methods for impurities? …
How Long to Equlibrate an HPLC Column?
MTS Helpdesk Question: How long does it take to equilibrate a column on an HPLC …
Services related to this content that we offer

The topic of forced degradation studies for the purposes of developing and validating stability indicating analytical methods is included in our courses, ‘How to Develop Stability Indicating HPLC Methods‘, and ‘Validation, Verification and Transfer of Methods for Pharmaceutical Analysis‘.

Visit our courses page for a full list of available courses and the schedule of currently available course dates.